Gingiva former (TA):
- for gingival forming
- 1.5 ° cone for a bacteria-proof connection and bone and papilla preservation
- Two-piece gingiva former in peek with separate screw available
- Abutments can be prepped
- Tightening Torque: 10 Ncm (2 mm Platform), 15 Ncm (3 mm & Short Platform)
|
 |
|
Provisional Abutments (PA):
- for making a temporary denture
- No clamp connection to the implant for easier manipulation of the structure
- can be customized individually
- Precisely fitting snap caps for closed impressions via transfer posts as well as direct abutment impression
- Gingivaformer cap as well as modeling caps as a temporary restoration are available Gingiva height (GH) of 1.5 mm
- Tightening Torque: 5 -10 Ncm (2 mm Platform) / 10-15 Ncm (3 mm & Short Platform)
|
 |
|
Preform Abutments (EAK):
- for the production of a cemented denture
- for incisor and canine area with 2 mm platform
- for all areas at 3 mm & Short Platform
- can be customized individually
- Precisely fitting snap caps for closed impressions via transfer posts as well as direct abutment impression
- Circular construction panel meets the highest aesthetic standards
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Tightening torque: 15 - 20 Ncm (2 mm Platform), 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
Anatomic Abutments (AA):
- for producing a cemented denture
- for incisor and canine area with 2 mm platform
- for all areas at 3 mm & Short Platform
- can be individually prepped
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Anatomical gingiva course of the abutment meets the highest aesthetic demands
- Tightening torque: 15 - 20 Ncm (2 mm Platform), 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
Freeform Abutments (FF):
- for producing an individual ligation structure with its own emergence profile, step or groove from a blank by manual processing (milling) of the technician
- should and must be prepared
- Anti-rotation of the crowns by groove and recess on the body
- for incisor. and canine area on 2 mm platform and all areas on 3 mm platform
- for crowns, bridges and telescopic structures
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Tightening torque: 15 - 20 Ncm (2 mm Platform) / 25 - 30 Ncm, for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
Castable Aufbauten (UCLA):
- For producing an individual alloy structure over the sprue with an individual emergence profile
- Construction for front and canine area with 2 mm platform and all areas with 3 mm platform
- For the production of a cemented denture
- Custom-fit burnout POM caps
- Abutment height provides a secure connection of the sprue construction
- Retentive depressions on the base structure with flattening as anti-rotation protection of the burn-out POM cap for individual wax modeling
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Circular construction panel meets the highest aesthetic standards
- Gingiva height (GH): 1.5 mm
- Suitable for telescopic crowns as well as individual modulation
- Tightening torque: 15 - 20 Ncm (2 mm Platform) / 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
Base Abutments (GAT):
- for the production of an individual zirconia Abutment by copying process with its own emergence profile
- The circular shape of the Abutment recesses enables ideal design freedom for the highest aesthetic demands
- for the production of a cemented denture
- for incisor and canine area with 2 mm platform
- for all areas at 3 mm & Short Platform
- Anti-rotation of the zirconium abutment by flattening the base structure
- abutment height ensures good bond strength of the milled zirconium structure
- Precisely fitting PEEK caps for data milling
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Tightening torque: 15 - 20 Ncm (2 mm Platform), 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
Scan/Glue Abutments (KSA):
- for the production of an individual zirconia abutment in the CAD-CAM process with its own emergence profile
- Scanbody or Scancap for digitizing the structure
- the circular shape of the abutment recesses enables ideal design freedom for the highest aesthetic demands
- for the production of a cemented denture
- for incisor and canine area with 2 mm platform
- for all areas at 3 mm & Short Platform
- Anti-rotation of the zirconium structure by cam on the base structure
- Construction height ensures good bond strength of the milled zirconium structure
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Tightening torque: 15 - 20 Ncm (2 mm Platform), 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
Cerec Abutments (CS):
- for the production of an individual zirconia abutment in the CAD-CAM process with its own emergence profile
- Precisely fitting Cerec scan cap for digitizing the abutment (Omni and Bluecam cap available)
- the circular shape of the abutment recesses enables ideal design freedom for the highest aesthetic demands
- for the production of hybrid abutments
- for incisor and canine area on 2 mm platform and all areas on 3 mm & Short Platform
- Anti-rotation of the zirconium structure by cam on the base structure
- Abutment height ensures good bond strength of the milled zirconium structure
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Tightening torque: 15 -20 Ncm (2 mm Platform) / 25-30 Ncm, for 3 mm Platform & 15-20 with Short Platform
|
 |
|
Emergence Abutments (EAZ):
- for the production of a screw-retained denture with its own emergence profile
- The circular shape of the abutment recesses enables ideal design freedom for the highest aesthetic demands
- for producing a screwed or cemented denture
- for incisor and canine area with 2 mm platform
- for all areas at 3 mm & Short Platform
- Anti-rotation of the crown due to the oval shape of the base structure
- Abutment height ensures good adhesion or screw strength
- 1.5 ° conical connection for maximum stability and bacteria tightness
- Tightening torque: 15 - 20 Ncm (2 mm Platform), 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
|
 |
|
RapidFix Abutments (VBA):
- for removable or screw-retained dentures for crowns, bridges and bars
- implant-supported RapidFix screwed supply
- 1.5 ° conical connection for a bacteria - proof connection and bone and papilla preservation
- Solid abutment: Two-piece construction with index
- Abutment screw with external hexagon for screwing in and internal thread for fixing the abutment with a short screw
- Analog and CAD suitable (data set integrated in K3Pro® library)
- Abutments can not be prepped
- Tightening torque: 15 - 20 Ncm (2 mm Platform), 25 - 30 Ncm for 3 mm Platform & 15 - 20 with Short Platform
-
|
 |
|
Multi Base Abutments (MB):
- for the removable or screw-retained dentures
- Implant-supported bar supply
- 1.5 ° conical connection for a bacteria-proof connection and bone and papilla preservation
- Solid abutment: two-piece design without index with screw for fixation of abutment and abutment with implant (one continuous screw)
- Analog and CAD suitable (data set integrated in K3Pro® library)
- Abutments can not be prepped
- Tightening torque: 15 - 20 Ncm (2 mm Platform) / 25 - 30 Ncm (3 mm & Short Platform)
|
 |
|
Finder Abutments (FLA):
- for removable dentures
- Mucosal and implant-supported locator supply
- 1.5 ° conical connection for a bacteria - proof connection and bone and papilla preservation
- Solid abutment: one-piece construction with integrated screw
- Abutments can not be prepped
- Tightening torque: 15 - 20 Ncm (2 mm Platform) / 20 - 30 Ncm (3 mm & Short Platform)
|
 |
|
O-Ring Abutments (KKA):
- for removable dentures
- Mucosa-supported O-ring supply
- 1.5 ° conical connection for a bacteria - proof connection and bone and papilla preservation
- Solid abutment: one-piece construction with integrated screw
- Abutments can not be prepped
- Tightening torque: 15 - 20 Ncm (2 mm Platform) / 20 - 30 Ncm (3 mm & Short Platform)
|
 |
|
Orthodontic Abutments (KFO):
- for orthodontic treatment
- for palate/ short implants
- secondary caps can be individually prepared
- circular abutment shoulder allows exact fit of the secondary cap
- Tightening torque: 15 - 20 Ncm (Short Platform)
|
|
|
|